KR2S Series Graphic Recorder

KR2S Series are advance touch screen display (Keyless) paperless Graphic Recorder with high performance and high operating function along with high visibility 5.7" VGA TFT Color LCD display. Universal input with high speed of sampling rate 100msec and high accuracy rating of ±0.1% realized. Measured data is stored into memory and support up to 8GB through USB and CF Card. As it can be monitored by a web browser display on several computers on intranet or internet, FTP transfer of data file and E-mail notification are also available.


Measuring points:
6 points, 12 points

Input types:
Universal
DC voltage --- ±13.8mV, ±27.6mV, ±69.0mV ±200mV, ±500mV, ±2V ±5V*, ±10V*, ±20V*, ±50V* (*with built-in voltage divider)
DC current --- With external shunt resistor (sold separately)
Thermocouple --- B, R, S, K, E, J, T, N, PtRh40-PtRh20, W-WRe26, WRe5-WRe26, PlatinelⅡ, NiMo-Ni, CR-AuFe, U, L
Resistance thermometer --- Pt100, JPt100, Pt-Co, Pt50

Accuracy ratings:
Refer to the table of measuring range and accuracy ratings

Reference junction compensation accuracy:
K, E, J, T, N, PlatinelⅡ--- ±0.5°C or less R, S, W-WRe26, WRe5-WRe26, NiMo-Ni, CR-AuFe,U, L --- ±1.0°C or less

Sampling rate:
Approximately 100ms for all points Burnout: Disconnection of input signal is detected on thermocouple and resistance thermometer input. UP/DOWN disable is selectable for each input

Scaling:
Range/scale is selectable when DC voltage/current is programmed.

Digital filter:
Programming FIR filter for each point (common to all points)

Allowable signal source resistance:
Thermocouple input (burnout disable)
DC voltage input (±2V or less) ---1kΩ or less
DC voltage input (±5V to ±50V) ---100Ω or less
Resistance thermometer --- Per wire 10Ω or less (same resistance for 3 wires)

Input resistance:
DC voltage, thermocouple input --- Approximately 1 MΩ

Maximum input voltage:
DC voltage input (±2V or less)
Thermocouple input (burnout disable) --- ±10VDC
DCvoltage input (±5V to ±50V) --- ±60VDC

Dielectric strength between channels:
1000V AC or more between each channel (High strength semiconductor relay used)