LE5000 Series Intelligent Recorders

LE5000 series are 250mm hybrid recorders with multi-range input. Innovative design high performance recorder provides high accuracy, ±0.05%; high speed scanning, 0.1second for 36 points and high speed recording, 1 line in 3 seconds. Simple operational keys and PC setting functions drastically improved usability of recording system.


High speed scanning at 36 points/sec and high-speed recording
Rapid changes of process data such as lab test results can be scanned simultaneously at 36 points/sec and recorded at about 1 line/ 3 sec. Data for each channel is displayed in 10 different colors which is user selectable.

High accuracy of 0.05%
The accuracy is ±0.05% and the resolution is 1μV or 0.1°C

Various industrial values can be measured at the same time with selectable ranges
With 35 thermocouple ranges and 8 DC voltage ranges, a total of 43 input ranges are provided which enables universal input and optional mixed input: current inputs are also possible.

Superior ease of operation
Operation keys are functionally designed for ease of use.

Engineering port is provided (USB)
A personal computer can be used as an engineering tool and parameter setting and data collecting is available.

Anti-noise countermeasures
High effective anti-noise countermeasures are taken; suppressive induced noise by 130 dB or more in the common mode while 50dB or more is achieved in the series mode. Effective countermeasures are taken against impulse noise.

Communication interfaces are available (Option)
RS-422A, RS-485 and Ethernet can be provided to meet various customers' needs.

Recording and calculation of data communication input (Option)
Data input by communications from a host can be recorded as analog and digital values at the same time with measuring data. Mathematical process of the data communications input from a host can be processed in parallel.

Measuring points: 12, 24 and 36 points

Input: Multi-channel data range
DC voltage --- ±10mV, ±20mV, ±40mV, ±80mV, ±1.25V, ±2.5 V, ±5 V, ±10V DC current --- Shunt resistor
(100Ω, 250Ω) needs to be mounted externally Thermocouple --- B, R, S, K, E, J, T, N, PtRh40-PtRh20,
NiMo-Ni,WRe5 – WRe26, W-WRe26, Platinel II, U, L
Resistance thermometer --- Pt 100, JPt 100

Range setting:
Input type and range are set with front keys

Scale setting:
The minimum and maximum values and unit are set for each point with front keys Setting range -30000 to 30000
Decimal points Optional setting

Indication accuracy:
Refer to items of measuring ranges, accuracy rating and display resolutions

Temperature drift:
0.1% FS/ 10°C

Measuring period:
0.1 sec/ channel

Reference junction compensation accuracy:
K, E, J, T, N, Platinel II --- ±0.5°C or less (0℃°C or more when measuring) R, S, WRe5-WRe26, NiMo-Ni, U, L --- ±1.0°C or less (Only when the ambient temperature is 23°C±5°C)

Input resolution:
Approx. 1/40000 (Standard range conversion)

Burnout:
Select with/ without burnout for each input

Allowable signal source resistance:
Thermocouple inputs, DC voltage input (10mV) --- 500Ω or less (without burnout) DC Voltage input (except 10mV) --- 100Ω or less Resistance thermometer inputs --- 10Ω or less/ line Three lines are common, Pt100, JPt100

Input resistance:
Thermocouple input, DC voltage input --- approx.1MΩ

Maximum input applied voltage:
±20V DC

Input correction:
Zero/span correction and shift correction for each channel

Maximum common mode voltage:
30V AC (support LVD) *250V AC at evaluation

Common mode rejection ratio:
130dB

Series mode rejection ratio:
50dB (Only when the peak value of noise is below standard range.)

Terminal board:
Detachable type, removable for wire connection